UVO-10k
Portable 10,000ppm UV Ozone Analyzer

 

The UVO-10k is a compact, handheld, precision UV ozone analyzer which can measure up to 10,000ppm ozone concentration. It has a precision and resolution of 10ppb. With its built-in long-life pump, 80,000 hour solid state UV LED light source and internal battery, it is the perfect instrument for the industrial, medical and environmental fields. It has a fast 1-minute warm-up which saves the user time. Its compact form makes it super portable and with its USB port, it can be charged, set up and logs downloaded.

Specifications:

  • Accurate and stable UV absorption method
  • Range: 0 to 10,000ppm
  • Accuracy: greater of 10ppb (1σ) [filter = 30s] or 2% of measurement
  • Resolution: 10ppb
  • Precision: greater of 10ppb (1σ) [filter = 30s] or 2% of measurement
  • Limit of detection: 20ppb (2σ) [filter = 30s]
  • Update rate: 1s
  • Response time: 2s (Filter = 2s), 20s (filter = 20s)
  • 10,000 hour pump
  • Solid state UV LED technology with 80,000 hour lifespan
  • Built-in temperature & pressure compensation
  • Operating temp (continuous): 0°C to +50°C
  • Operating humidity (continuous non-condensing): 0% to 95%
  • Humidity equalization technology
  • Ozone units: ppm, mg/m³
  • Temp units: K, °C, °F
  • Pressure units: kPa, mbar, torr
  • Flow rate: 0.5lpm (nominal); 0.2lpm (min.)
  • Averaging modes: 2s-120s, 1min, 5min, 15min, 30min, 1h
  • On-board data logger memory: 32000 lines of data
  • Backlit LCD display
  • Rechargeable battery operation: 8 hours (backlight off)
  • Programmable alarm levels with alarm buzzer
  • Warm up time: 1 min
  • USB charging
  • USB data download
  • Size: 160mm x 100mm x 51mm

Applications:

  • Industrial hygiene monitoring where ozone may be present in the workplace.
  • Research and quality checking of ozone emitting equipment.
  • Monitoring high ozone concentrations such as process off-gassing.
  • Long-term monitoring at remote locations where power is highly limited
  • Monitoring and control of ozone in industrial settings
  • Monitoring of exposure to individuals in the workplace